Set Sample Locations In The Batch Editor; Select Vial Positions Using The Locations Tab (Optional) - AB Sciex QTRAP 4500 LC/MS/MS system System User's Manual

Qtrap 4500 series
Table of Contents

Advertisement

Item Description
2
3
3. Click Acquire > Start Sample.
Note: It is recommended that the user runs the sample again in the event
of an abnormal termination during sample acquisition.

Set Sample Locations in the Batch Editor

If an autosampler is used in the acquisition method, then the vial positions of the samples must
be defined in the acquisition batch. Define the location in the Sample tab or in the Locations tab.
For more information about creating batches, refer to
page
64.
Note: Depending on the autosampler being used, it may not be necessary to
type details in additional columns.
1. In the Sample tab, in the Set list, select the set.
2. For each sample in the set, do the following if applicable:
In the Rack Code column, select the rack type.
In the Rack Position column, select the position of the rack in the
autosampler.
In the Plate Code column, select the plate type.
In the Plate Position column, select the position of the plate on the rack.
In the Vial Position column, type the position of the vial in the plate or tray.
3. Save the file.
Select Vial Positions using the Locations Tab
(Optional)
1. In the Batch Editor, click the Locations tab.
System User Guide
RUO-IDV-05-0264-A
Queue status should be Stand By.
Queue Server should be in Normal mode. Refer to
States on page
71.
Operating Instructions — Batches
Queue
Add Sets and Samples to a Batch on
4500 Series of Instruments
67 of 138