Figure 11-1 Create Quantitation Set - Select Settings & Query Page
6. Click Next.
Figure 11-2 Create Default Query Page
7. In the Maximum Allowed Accuracy Variation for QCs (%) table in the Max.
Variation column, type the maximum allowable percent of variation for each QC (for
example, 5 is ± 5%) in the same row as the corresponding concentration. If the
concentrations were not specified during acquisition, they are not shown here. In that
case, type them in the Concentration column.
8. In the Maximum Allowed Accuracy Variation for Standards (%) table, in the Max.
Variation column, type the maximum allowable percent of variation for each
System User Guide
RUO-IDV-05-0264-A
Operating Instructions — Analyze and Process Quantitative Data
4500 Series of Instruments
99 of 138