Computer and Network Requirements

Return to checklist.

Acquisition Computer

The acquisition computer and monitor are provided with the mass spectrometer. It controls the mass spectrometer.
This computer should be used exclusively for data acquisition.
• Provide a table for the acquisition computer and monitor, within 3 m (120 inches) of the mass spectrometer.
• Provide a computer name and password that comply with these guidelines:
• The computer name does not have any spaces.
• The password for the computer is not blank.
Note: If the acquisition computer will be connected to the LAN, then this must be a network computer
name and password. Refer to
CAUTION: Potential System Damage: Other than security software, do not install
additional software on the acquisition computer. Changes to the configured software
could void the system warranty and cause the system to stop operating.
• If security software is required, then make sure that an IT specialist is available to install the security software
(anti-virus, anti-malware, or backup software) while the FSE is present.

Printer Requirements

Return to checklist.
The system can be connected to a network or dedicated printer.
Note: To use a network printer, connect the acquisition computer to the company network.
• Make sure that a printer and its associated print drivers are available before the scheduled installation.

LAN Connection (Optional)

Return to checklist.
If you want to connect the acquisition computer to the network, then:
• Make sure that an active, tested LAN connection is in place before the scheduled installation date.
Site Planning Guide
RUO-IDV-01-0563-A | D5040008 A
LAN Connection (Optional) on page
Site Requirements
27.
AB SCIEX 4000 Series of Instruments