Set The Gc Column Parameters - Agilent Technologies G1701EA Familiarization Manual

Msd productivity chemstation
Hide thumbs Also See for G1701EA:
Table of Contents

Advertisement

Set the GC column parameters

Figure 10 GC columns parameters
Agilent G1701EA MSD Productivity ChemStation Familiarization Guide
Select the Columns button,
1
are displayed.
Check the Column information in the Selection list.
2
Column: 19091S-433 (HP-5ms 30 m x 250 µm x 0.25 µm)
In: front or back (split/splitless inlet position)
Out: Vacuum
Mark the Control Mode checkbox.
3
In the Flow Setpoint field, enter 1.0 mL/min. The Pressure,
4
Average Velocity, and Holdup Time Setpoints will be calculated
and displayed in the corresponding fields.
In the dropdown list, select Constant Flow.
5
In the Post Run field, enter 1.0 mL/min.
6
Select Apply to download these settings to the GC.
7
Create a Method for Qualitative Analysis
. The Column parameters
3
35