Shut Down the GC
Agilent G1701EA MSD Productivity ChemStation Familiarization Guide
Do not turn off the MS at this time if you are first cooling
down the instrument. The Instrument Control window will
close when a configured instrument is powered off.
Select Close.
6
In Instrument Control, load the GC cool down method.
1
Access the Edit GC Parameters window.
2
Right mouse click in the right panel and select Download
3
Method to GC from the context menu. A confirmation
message is displayed.
Select OK to close the message and return to the GC Edit
4
Parameters window.
Close the Edit GC Parameters window and exit the
5
ChemStation.
When the GC enters the Ready state turn off the power to
6
the GC and the MS.
Turn off the carrier gas.
7
Power off the PC and all peripheral equipment.
8
12
Shut Down the System
143