3
Create a Method for Qualitative Analysis
Figure 6
Column configuration tab
30
Select the Columns tab to display the columns configuration
3
parameters. The HP-5ms checkout column supplied with the
MS should be listed under Column.
If a different Column is configured to the inlet location you
4
are using or is attached to the MS, select it and click Remove.
If the HP-5ms is not listed under Column, click the Inventory
5
button and add it to inventory before listing it here. See
"Add a column to ChemStation local inventory"
If required, use the up and down arrow keys to put the
6
HP-5ms column in the 1 position.
For the Inlet pressure for this column, select the Front or Back
7
Inlet from the dropdown.
For the column Outlet pressure select Vacuum for the MS
8
For the column Heated By select Oven from the dropdown.
9
Agilent G1701EA MSD Productivity ChemStation Familiarization Guide
on page 48.