12
Shut Down the System

Shut Down the MS

142
Select View > Tune and Vacuum Control....
1
Select Vacuum > Vent... A confirmation message appears.
2
Select OK.
3
The Vent Cycle dialog status window opens and remains open
4
until the vent is complete. You can close the dialog box by
selecting Exit, however, the process continues. To reopen the
Vent Cycle status window, select View > Vacuum Status.
Select OK to close the dialog box.
5
Agilent G1701EA MSD Productivity ChemStation Familiarization Guide