Display Extracted Ion Chromatograms (Eic)S - Agilent Technologies G1701EA Familiarization Manual

Msd productivity chemstation
Hide thumbs Also See for G1701EA:
Table of Contents

Advertisement

5
Qualitative Data Analysis

Display Extracted Ion Chromatograms (EIC)s

76
Select the Ion Chromatograms button
1
Chromatograms dialog box opens.
In the Time Range fields, enter the range you wish to extract.
2
The complete time range of the data file is initially
displayed. You can specify a shorter time range by entering
the appropriate starting and ending values.
In the Ions area, enter the ion masses of interest. You can
3
specify up to six ions.
In the Use m/z range from fields, enter the range of interest.
4
The default m/z range for each ion is -0.3 to +0.7 of the ion
mass specified. You can change the range by entering the
appropriate starting and ending values.
Select OK. A window opens displaying a chromatogram for
5
each ion.
Agilent G1701EA MSD Productivity ChemStation Familiarization Guide
. The Extracted Ion