Agilent G1701EA MSD Productivity ChemStation Familiarization Guide
In the scan window, position the bulls eye cursor on the
5
target ion (154) and click both mouse buttons
simultaneously. The m/z is displayed for the Target.
Q1 is selected in the Quant Setup dialog box.
In the scan window, position the cursor on the first qualifier
6
ion (153) and click both mouse buttons simultaneously. The
m/z is added to the Q1 field and the ratio is calculated and
added to the Ratio field.
To clear an incorrect ion selection, select the radio button
for that ion. Next, simultaneously click both mouse buttons
with the cursor positioned on an area not containing an ion.
Select Save to add the biphenyl peak to the database and
7
clear the Quant Setup dialog box.
8
Add the remaining compounds using the target and qualifier
ions identified in qualitative analysis.
Table 6
Target and qualifier ion selections
Compound
Biphenyl
Dodecane
Chlorobiphenyl
Methyl pamitate
When all compounds are added, select Exit to return to the
9
Edit Compounds dialog box.
Review the compound list. If any corrections need to be
10
made, double-click on the compound and reenter the
information in the Quant Setup dialog box.
Set Up a Quantitation Database
Target Ion
Qualifier Ion
154
153
170
85
188
152
270
87
8
Dwell time
60
60
60
60
113