To Set The Gc/Msd Interface Temperature From The Chemstation - Agilent Technologies 5975T LTM GC Operation Manual

Table of Contents

Advertisement

To Set the GC/MSD Interface Temperature from the ChemStation

Procedure
1 Select View>Instrument Control.
2 Select Instrument>GC Edit Parameters.
3 Click the Aux icon to edit the interface temperature
Figure 20
Setting the interface temperature
4 Check the heater On and type the setpoint in the Value °C column.
5975T LTM GC/MSD Operation Manual
Operating in Electron Impact (EI) Mode
(Figure
3
20).
81