3
Create a Method for Qualitative Analysis

Set the GC inlet parameters

36
Select the Inlets button,
1
displayed.
Select the Front or Back tab, depending on your hardware
2
configuration.
Mark the Heater checkbox and enter 250 °C in the
3
corresponding Setpoint field.
Mark the Pressure checkbox. The psi in the corresponding
4
Setpoint field is automatically set when the column flow rate
is set.
Mark the Septum Purge Flow checkbox and enter 3 mL/min in
5
the corresponding Setpoint field.
From the Septum Purge Flow Mode drop down list, select
6
Standard.
In the Gas Saver area:
7
Mark the On checkbox.
a
b
In the field below, enter
In the After field, enter
c
In the Mode area:
8
From the Mode drop down list, select Splitless.
a
In the Purge Flow to Split vent area:
9
a
In the field, enter
In the Start Time field, enter
b
Select Apply.
10
Agilent G1701EA MSD Productivity ChemStation Familiarization Guide
. The Inlet parameters are
20 mL/min
.
2 min
.
50 mL/min
.
1
.