2
Tune the MS
Evaluate the Autotune Results
Figure 3
Passing system verification tune report
22
Select View > Instrument Control.
1
Select Checkout > Evaluate Tune. The system compares your
2
tune parameter results to preset acceptable results and
displays the System Verification report. See
Review the report. Criteria marked as OK are functioning
3
correctly. If all criteria are marked OK, Tune portion of System
Verification passed is printed on the last line of the report. See
Figure
3.
If one or more criteria do not pass verification, the incorrect
behavior and suggested corrective actions are described. See
Figure 4
on page 23 where the report shows a high ratio of mass
18 to 69. This report warns of a high amount of water in the
system and a corrective action to be taken.