Take a Snapshot

Figure 31 The TIC of the snapshot data file
Agilent G1701EA MSD Productivity ChemStation Familiarization Guide
Snapshot is useful when a compound of interest elutes early
during a long run and you want to analyze that compound
immediately. The system creates a snapshot data file with data
that has been acquired up to the time the Snapshot is taken.
During the run select View > Data Analysis to open the data
1
analysis view.
Select File > Take Snapshot. The data analysis windows opens
2
displaying the TIC obtained for the run up to this point in
time.
Observe the location of the snapshot data file in the
navigation pane. It is placed in the data directory specified
for the run under the snapshot subdirectory and given the
same name as the data file specified for the sample.
Analyze the compound of interest.
3
Exit data analysis and return to the Instrument Control view.
4
4
Run the Scan Method