Create The Cool Down Method - Agilent Technologies G1701EA Familiarization Manual

Msd productivity chemstation
Hide thumbs Also See for G1701EA:
Table of Contents

Advertisement

11
Create a Cool Down Method

Create the Cool Down Method

138
Select View > Instrument Control.
1
Select the GC Edit Parameters button,
2
Parameters window opens.
Select the Oven button,
3
displayed.
In the Oven Ramp table, clear the Rate and Value entries.
4
Select the Inlets button,
5
displayed.
Select the front or back tab, depending on your hardware
6
configuration.
Mark the Heater check box and enter
7
corresponding field.
Mark the Pressure check box. Column flow must be
8
maintained to prevent damage to the column when hot.
Select the AUX button,
9
Clear the On check box for the Aux 2 Heater.
10
Agilent G1701EA MSD Productivity ChemStation Familiarization Guide