Environmental chambers TC30 and TC40
Setting the desired temperature
The desired temperature is set by using the two arrow keys (
increase;
Pressing the key briefly changes the temperature in one-tenth of a degree
steps.
Fast adjustment is carried out by pressing the key longer.
No functions are assigned to the two keys
11.8.3

Setting the thermostat

With the TC40 the thermostat is used to mediate between room temperature and
target temperature.
!
The given temperature limits of 160°C or -30°C can only be achieved when the
Peltier element is supported by the thermostat.
As a rule of thumb the temperature difference between the measuring
temperature and the thermostat liquid should not exceed 30°C when cooling
down and 70°C when heating up. However, greater temperature differences can
be used in the temperature range between –10 and 120°C.
The following table assumes that water is used as the thermostat liquid:
Measuring temperature range
-30°C to 10°C
-10°C to 120°C
50°C to 160°C
SPA C SPA C SPAC SPAC
= temperature decrease).
and
Recommended thermostat setting
2°C
20°C
85°C
page 79