Appendix
Setting the boards
The first module to be plugged receives the address 4, the second and third the
addresses 5 and 6 in ascending order.
Turn the screw with the help of a screw driver so that the slit points to the
required number.
Port assignment for multi-dosing units DS3222-3228
Both the syringe revolver and the syringes drivers are connected to the electronic
module(s):
The connections of the syringe revolver always occupy ports 1 and 2 of the first
electronic module (address 4).
The syringe drivers occupy the further ports in a special order that depends on
the number of drivers. The plugs coming from the syringe drives are marked
with 1 to 4 dots. The left syringe driver unit is assigned to syringe numbers 1-4,
the right-hand side unit to numbers 5-8.
Vc)
Mounting
Safety instructions
Switch off and unplug the DSA100 before starting with the installation.
Damage by electrostatic charge
Touching the electronic module when the body is charged electrostatically can
damage electronic components.
Before touching the electronics board discharge yourself, for example by
touching a metallic part of the panel.
'
addresses
page 107