12. Place the vessel in the position that corresponds to syringe 1.
The storage bottles delivered by KRÜSS stick to the bottom of the rack
magnetically.
13. Switch the valve used back to channel the liquid from the syringe to the dosing
tube ( see step 2).
14. Repeat the procedure for the other syringe(s).
SPA C SPA C SPAC SPAC
15. Screw the needles into the needle carriers.
The thin needles are intended for sessile drops, the thicker ones for pendant
drops (surface tension measurements).
SPA C SPA C SPAC SPAC
10.3

Adjusting the dosing unit

The dosing unit must be adjusted with respect to the camera so that the middle
horizontal needle position is in the middle of the recorded image.
1. Start the DSA software.
The camera button
image). Click on the button if necessary.
2. Select the tab
3. Select Syringe 1 (S1) in the drop-down menu in the upper left.
If a different syringe had been selected previously the syringe selector will
rotate to Position 1.
4. Select the tab
in the toolbar must be shown as being pressed (live
Dosing
from the
Device Control
DSA1
Needle Pos
(DSA1) respectively
DSA1
Multi-dosing unit DS3226-3228
panel.
DSA3/DSA4
N-Pos
(DSA3/DSA4).
DSA3/DSA4