page 110
6. Connect the two plugs coming from the connection of the dosing unit to the
two ports 1 and 2 of the installed electronics board in an arbitrary order.
7. Only for multi-dosing unit DS3222-3228: Connect the other ports of all
electronic modules as described in the table on page 105.
!
Pay attention to the assignment of the plugs 1-4 and 5-8 coming from the
syringe drives to the ports 1-4 of the modules to be connected.
8. Return the supporting frames and fix them.
!
Take care that the edge of the board lies in the gaps of the frames.
9. Reassemble the covers removed in steps 2-3.
SPA C SPA C SPAC SPAC