kruss DSA100 Installation And Operation Manual page 19

Drop shape analyzer
Table of Contents

Advertisement

Computer system
Step 4: Firmware check / frame grabber update
!
If a frame grabber update is necessary (see subsequent instructions), this step
must not be omitted. The frame grabber can be damaged when working with a
dated firmware.
1. Start the program
2. If a frame grabber update is necessary this is stated in a program message.
Confirm this message with OK. The frame grabber gets updated automatically.
If no message appears, no update is necessary. In this case just terminate
the program Intellicam.
3. After the upgrade the computer switches off automatically. Reboot the
computer.
Step 5: Reducing the hardware acceleration
1. Start
the
ImagingTools.
2. Select the
3. Move the slider two tick marks to the left. The text
mono images
4. Click OK.
SPA C SPA C SPAC SPAC
Intellicam
from the start menu folder
program
MilConfig
Video Hardware Acceleration
must appear in the lower left of the scale.
Matrox
from the start
menu folder
tab.
Disable DDraw Overlay for
page 17
Imaging.
Matrox