Appendix
7. Lift the syringe driver unit to be assembled into the compartment and lead the
connecting cables through the gap that was previously covered.
!
If two different syringe driver units are to be assembled please pay attention
to the correct assignment to the left and right-hand compartment.
8. Fix the syringe driver unit (in example picture: unit with two drivers) with the
four previously removed hexagon socket screws using the corresponding holes.
9. Push through the rubber stoppers with a pointed object.
page 99