Appendix
13. Take one fitting and one ferrule (see image) from the accessories case for each
dosing tube. Thread the free end of the tube in the compartment through a
fitting and put a ferrule on the tip of the tube.
Fitting
14. Screw off the protective caps from the valve blocks and screw on the fitting with
tube and ferrule.
!
Screwing the fitting too tight can damage the ferrule and squeeze off the
tube. Do not use tools; the fitting should be easily unscrewed manually.
15. Put the rack for the test liquid bottles into the chamber(s).
Leave the DSA100 housing open in order to install the electronic module(s)
EM3201.
SPA C SPA C SPAC SPAC
page 101
Ferrule