Addresses and mounting of the electronic module
V
EM3201
Depending on the dosing unit used, one, two or three electronics board must be
addressed, mounted and connected.
Va)
Overview
The following table contains an overview for the electronic modules EM3201 to be
installed for the different dosing units.
Dosing unit
DS3210 (Single
Direct Dosing)
DS3222 (2 syringes)
DS3223 (3 syringes)
DS3224 (4 syringes)
DS3225 (5 syringes)
DS3226 (6 syringes)
Number of
Addresses
EM3200
1
1
2
4, 5
2
4, 5
2
4, 5
2
4, 5
Slot assignment
4
1,2
Module 1 (Adr. 4)
4
1,2: syringe revolver
3,4: syringe drivers 1, 2
Module 1 (Adr. 4)
1,2: syringe revolver
Module 2 (Adr. 5)
1,2,3: syringe drivers 1-3
Module 1 (Adr. 4)
1,2: syringe revolver
Module 2 (Adr. 5)
1,2,3,4: syringe drivers 1-
4
Module 1 (Adr. 4)
1,2: syringe revolver
3,4: syringe drivers 1, 2
Module 2 (Adr. 5)
1,2,3: syringe drivers 3-5
Module 1 (Adr. 4)
1,2: syringe revolver
3,4: syringe drivers 1, 2
Module 2 (Adr. 5)
1,2,3,4: syringe drivers 3-
6