8. Subsequent retrieval of the last melting point

8. Subsequent retrieval of the last melting point
If you missed the time for reading the melting point and
the device has returned to the starting display, the last
melting point can be displayed once more by pressing
the printing key.
The display will appear for about 5 sec. After that the
device will return to the starting display. If the printing
key is activated once more within the 5 s, the print-out
will be created with the attached printer (if available).
This can be repeated as often as you want.
The term "auto" will not be shown on the display, since
no melting point analysis can be started in this case
inserting a capillary tube does not have any effect.
If an internal error value is determined as melting point,
the melting point analysis was prematurely aborted by
pressing the arrow keys (clear text display of the error
for 5 s).
The melting point could not be determined due to a
time-out (10 min.), i.e. the melting point could be
10 °C higher than the selected starting temperature
(clear text display of the error for 5 s).
26
actual temperature
25.8 °C
melting point
121.9 °C
abort,
no melting point
9. Setting of date, time and language / 10. Technical data

9. Setting of date, time and language

After pressing the time-set-key on the right hand side of
the display, the set-up menu appears.
The settings will be made in the following order:
return with:
A flashing cursor will appear on the display. With each
activation of the time- / set-key the cursor will jump to
the next item. The value can be changed resp. set with
the arrow keys. When setting the date you should start
with year, then month and at least day and week-day.
When the month is set, the day always shows 1.
Hour and minute are set separately.
When all settings are completed, the menu is left and
all settings are saved by pressing the printing key. The
settings will also remain after turning off the device.
The device includes a battery. This is why the date and
time are also updated in the off mode for about 1 year
without turning on the device.
Calibration parameters as well as language are stored
in EEPROM and therefore remain for a longer period.
min
h
weekday
year
month
day
language
print
english12:38
Mo 31.01.2009