4. Select the syringes one after another (except the one marked with M) and carry
out the menu order
(DSA3/DSA4) for each syringe. The syringe drivers are moved into their start
positions and initialized.
SPA C SPA C SPAC SPAC
10.2

Inserting the syringes

SPA C SPA C SPAC SPAC
1. Select the first syringe in the DSA1/DSA3/DSA4 software ( Chapter 10.1).
2. In order to remove air from the refill tube the dosing valve must be switched to
channel the liquid from the syringe to the refill tube.
For DSA1: Select the tab sheet
the
Valve Position for Dosing
Position for Refill.
For DSA3/DSA4: Open the drop-down menu for syringe selection. Select the
menu item
opened a dot indicates that the valve is switched to the refill tube.
Reset Position
DSA1
Tools
Toggle Switch Syringe
(DSA1) respectively
DSA3/4
Refill
from the
Device Control
button. The button label changes to Valve
Valve. The next time the menu is
page 63