Adjusting The Dosing Unit - kruss DSA100 Installation And Operation Manual

Drop shape analyzer
Table of Contents

Advertisement

page 60
2. Insert the syringe in the syringe adapter.
To do this, push the syringe plunger upwards and insert the plunger.
Clip the glass syringe into the clamping springs so that the top edge of the
syringe lies between the top clamping spring and the end spring.
3. Fix the plunger of the syringe in the clamp with the help of the screw which was
loosened in Step 1.
SPA C SPA C SPAC SPAC
9.3

Adjusting the dosing unit

The dosing unit must be adjusted with respect to the camera so that the middle
horizontal needle position is in the middle of the recorded image.
1. Start the DSA software.
The camera button
image). Click on the button if necessary.
SPA C SPA C SPAC SPAC
2. Select the tab
SPA C SPA C SPAC SPAC
3. Use the vertical slide controller or the <↓> key to move the needle to a position
between the camera and the light source.
SPA C SPA C SPAC SPAC
in the toolbar must be shown as being pressed (live
Needle Pos
(DSA1) respectively
DSA1
Software-controlled single-Dosing unit DS3210
N-Pos
(DSA3/DSA4).
DSA3/DSA4