4. Set the illumination, zoom and focus for the needle image. The needle should
be in focus and take about one tenth of the image width.
For software-controlled optics, this is carried out in the tab
respectively
5. Use the setscrew on the right-hand side of the dosing unit to bring the needle
in the middle position in the video image.
SPA C SPA C SPAC SPAC
9.4

Restricted software options

Some functions of the DSA3/DSA4 software are restricted to multi-dosing units and
cannot be used for DS3210.
Due to the simplified syringe control of the DS3210, no information about the
current syringe driver position is transferred to the software. Therefore, the
filling level display and the Auto-Refill function do not work for single-dosing.
The Auto-Delivery function for automatic surface detection is also blocked.
For further information please consult the DSA3/DSA4 software manual.
SPA C SPA C SPAC SPAC
O-Con
(DSA3/DSA4) ( software manual).
page 61