6 INSTALLATION

6.1 Siting Requirements

Notes.
• Mount in a location free from excessive vibration.
• Mount away from harmful vapours and/or dripping
fluids.
• Where possible, mount the analyzer at eye level to
allow an unrestricted view of the front panel displays
and controls.
Maximum Distance
cell constant <0.1
cell constant 0.1
A – Maximum Distance Between Analyzer and Sensor
–20C
(–4F)
B – Within Temperature Limits
NEMA 4X
C – Within Environmental Limits
Fig. 6.1 Siting Requirements
50
Conductivity
Sensor
50m (137.5 ft) –
100m (275 ft) –
65C
(149F)
Max.
Min.
IP66