...5.5 Configure Outputs
Off
-----
On
Hold
mA
AO1: Default O/P
Config. Output 1
Off or Hold
On
12. 0 0
mA
AO1: Default Val
Config. Output 1
Default Output
Select the system reaction to failure:
Off
– Ignore failure and continue operation.
On
– Stop on failure. This drives the analog output to the level set in the
frame below.
Hold
– Hold the analog output at the value prior to the failure.
Default Value
The level to which the analog output is driven if a failure occurs.
Set the value between 0.00 and 22.00mA.
Output 2 configuration (and Outputs 3 and 4 if option board fitted and analog
Config. Output 2
features enabled – see Section 7.3) is identical to Output 1 configuration.
Option board fitted and analog features enabled (Section 7.3) – see Section 5.7.
CONFIG. CLOCK
Option board fitted and Serial Communications feature enabled (Section 7.3) –
CONFIG. SERIAL
see Supplementary Manual PROFIBUS Datalink Description (IM/PROBUS).
Single input analyzer and option board not fitted – see Section 5.8.
CONFIG. CONTROL
Dual input analyzer and option board not fitted – see Section 5.9.
CONFIG. SECURITY
5 PROGRAMMING...
Default Val
37