ABB AX410 User Manual page 27

Single and dual input analyzers for low level conductivity
Hide thumbs Also See for AX410:
Table of Contents

Advertisement

...5.3 Configure Sensors
A: T.Comp Range
set to
Hi TC
Acid
Neutrl
UPW
-----
Base
A: Temp.Comp.
High Range Temperature Compensation
Select the type of high range (0 to 200C [32 to 392F]) temperature compensation
required:
Base*
– Temperature compensation based on the temperature coefficient of pure
water containing trace alkalis.
UPW*
– Temperature compensation based on the temperature coefficient of pure
water. Source data is based on International Standard IEC 60746-3.
Also enables manual entry of a temperature coefficient (see
frame on page 26) for applications where pure water contains an unknown
impurity; in which case the temperature coefficient must be calculated –
see Appendix A1.1.
Neutrl*
– Temperature compensation based on the temperature coefficient of pure
water containing trace neutral salts.
Acid*
– Temperature compensation based on the temperature coefficient of pure
water containing trace acids.
Examples
• Cation exchanger in-bed and outlet applications.
• Degassed cation/conductivity applications.
* Applicable only on conductivities up to 10S cm
Continued on next page.
A: Temp.Sensor
5 PROGRAMMING...
Temp.Coeff.
–1
25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax411Ax450Ax455Ax456Ax413Ax416 ... Show all

Table of Contents