A: T.Comp Range
Temperature Compensation Range
Select a temperature compensation range to suit the sample temperature:
Off
– Raw conductivity measurement without temperature compensation.
Examples
• Water for injection (WFI) for US Pharmacopoeia (USP) applications.
• Purified water for USP applications.
– Temperature compensation for sample temperatures in the range 0 to
Lo TC
100C (0 to 212F). This setting is suitable for the majority of applications.
Hi TC
– Temperature compensation for sample temperatures in the range 0 to
200C (0 to 392F). This setting is used only for special high temperature
applications.