...5 PROGRAMMING
...5.4 Configure Alarms
set
A1: Type
to
USP645
0. 0 0
uS/cm
A1: USP Offset
not
A1: Type
set to
USP645
100. 0
uS/cm
A1: Setpoint
0. 0
%
A1: Hysteresis
0
Secs
A1: Delay
Config. Alarm 1
32
USP Offset
Enables the USP645 alarm set point value to be adjusted for increased process protection,
i.e. the USP alarm set point value in Table 5.3 is offset by the amount entered (table value
– offset value) to enable the alarm to be triggered early.
Note. The
parameter is available only on AX450 and AX455 analyzers and only
USP Offset
if
is set to
A: Cond.Units
USP645
Alarm 1 Set Point
Set the alarm set point to a value within the input range being displayed – see Table 5.2
(page 22).
Alarm 1 Hysteresis
A differential set point can be defined between 0 and 5% of the alarm set point value. Set
the required hysteresis in 0.1% increments.
See also Figs. 5.1 to 5.5 (page 33).
Alarm 1 Delay
If an alarm condition occurs, activation of the relays and LEDs can be delayed for a
specified time period. If the alarm clears within the period, the alarm is not activated.
Set the required delay, in the range 0 to 60 seconds in 1 second increments.
See also Figs. 5.1 to 5.5 (page 33).
Alarms 2 and 3 configuration (and Alarms 4 and 5 if option board fitted and
Config. Alarm 2
analog features enabled – see Section 7.3) is identical to Alarm 1.
See Section 5.5.
CONFIG. OUTPUTS
(Section 5.3) and
is set to
A1: Type
.
USP645