...7.3 Factory Settings
5000
uS/cm
5000
Calib
A:Res.Span(20R)
5014
uS/cm
5014
Calib
A: Self Checking
100. 0
Ohms
100. 0
Calib
A:T.Zero
(100R)
150. 0
Ohms
150. 0
Calib
A:T.Span
(150R)
1000
Ohms
1000
Calib
A:T.Zero
(1K0)
1500
Ohms
1500
Calib
A:T.Span
(1K5)
Cal. Sensor A
No
-----
Yes
A: Abort Cal.
Resistance Span (20R)
Set the cell simulator to 20.
The display advances automatically to the next step once a stable and valid value is
recorded.
Self Checking
The analyzer calibrates the internal reference resistance automatically to compensate for
changes in ambient temperatures.
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (100R)
Set the temperature simulator to 100.
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (150R)
Set the temperature simulator to 150.
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (1k0)
Set the temperature simulator to 1k.
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (1k5)
Set the temperature simulator to 1.5k.
The display returns automatically to
Abort Calibration
Select Yes or No
selected:
Yes
completion of
before
A:T.Zero (100R)
A:T.Zero (100R)
completion of
after
CAL. SENSOR A
Calibrate Sensor A
selected – calibration continues from the point at which the
No
pressed.
once a stable and valid value is recorded.
Cal. Sensor A
frame – calibration advances to
A: Self Checking
and continues.
frame – the display returns to the
A: Self Checking
page.
7 CALIBRATION...