ABB AX410 User Manual page 28

Single and dual input analyzers for low level conductivity
Hide thumbs Also See for AX410:
Table of Contents

Advertisement

...5 PROGRAMMING
...5.3 Configure Sensors
Pt100
-----
Pt1000
A: Temp.Sensor
2. 0 0
%/C
A: Temp.Coeff.
0. 7 5
ppm/uS
A: TDS Factor
mg/kg
mg/l
-----
ppm
A: TDS Units
Yes
-----
No
A: Enable Cals.
Config. Sensor A
26
Temperature Sensor
Select the type of temperature sensor used, Pt100 or Pt1000.
Temperature Coefficient
Notes.
Displayed only if
T.Comp Range
or
is set to
T.Comp Range
If
is set to
A: Cond Units
USP645
temperature coefficient is fixed automatically at 2.00%/C.
Enter the temperature coefficient ( x 100) of the solution (0.01 to 5.0%/C). If unknown,
the temperature coefficient () of the solution must be calculated – see Appendix A1.1.
If the value has not yet been calculated, set it to 2%/C provisionally.
TDS Factor
Note. Displayed only if
A: Cond.Units
The TDS factor must be programmed to suit the particular application – see Appendix A2.
Enter the required TDS factor between 0.4 and 0.8.
For salinity applications, set the TDS factor to 0.6.
TDS Units
Note. Displayed only if
A: Cond.Units
Select the TDS units (ppm, mg/l or mg/kg).
Enable Calibration
Select
to enable sensor calibration – see Section 4.1.
Yes
If
is selected the calibration menu, pages and frames for the relevant sensor are not
No
displayed.
Sensor B configuration (dual input analyzers only) is identical to Sensor A
Config. Sensor B
configuration.
Single input analyzers only – return to main menu.
CONFIG. SENSORS
See Section 5.4.
CONFIG. ALARMS
is set to
and
Lo TC
Temp.Comp.
and
is set to
Hi TC
Temp.Comp.
(AX450 and AX455 analyzers only – see page 22), the
is set to
– see page 22.
TDS
is set to
– see page 22.
TDS
is set to
or
Linear
UPW
– see pages 23 to 25.