ABB AX410 User Manual page 68

Single and dual input analyzers for low level conductivity
Hide thumbs Also See for AX410:
Table of Contents

Advertisement

...7 CALIBRATION
...7.3 Factory Settings
-----
Cal. Output 1
16000
O1: Adjust 4mA
7200
O1: Adjust 20mA
Cal. Output 1
-----
Cal. Output 2
7200
O1: Adjust 20mA
Cal. Output 2
66
Calibrate Output 1
Note. When adjusting the 4 and 20mA output values, the display reading is unimportant
and is used only to indicate that the output is changing when the
pressed.
See below.
Cal. Output 2
Adjust 4mA
Use the
and
keys to set the milliammeter reading to 4mA.
Note. The analog output range selected in
the reading.
Adjust 20mA
Use the
and
keys to set the milliammeter reading to 20mA.
Note. The analog output range selected in
the reading.
See below.
Cal. Output 2
Conductivity
Operating Page – see Section 2.3.
Dual Cond.
Calibrate Output 2
Note. Output 2 calibration is identical to Output 1 calibration.
Option board fitted and analog features enabled – continued on next page.
Cal. Output 3
Option board fitted, additional features disabled – continued on next page.
Option Board
Option board not fitted – continued on next page.
Alter Fact. Code
Option board fitted and analog features enabled – continued on next page.
Cal. Output 3
Option board fitted, additional features disabled – continued on next page.
Option Board
Option board not fitted – continued on next page.
Alter Fact. Code
Conductivity
Operating Page – see Section 2.3.
Dual Cond.
(Section 5.5) does not affect
Configure Outputs
(Section 5.5) does not affect
Configure Outputs
and
keys are