...7 CALIBRATION
...7.3 Factory Settings
0. 0
uS/cm
0. 0
Calib
A:Res.Zero (O/C)
500. 0
uS/cm
500. 0
Calib
A:Res.Span (2K0)
0
uS/cm
0
Calib
A:Res.Zero (O/C)
64
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.
Note. The upper 7-segment display shows the measured conductivity. Once the signal is
within range the lower 7-segment display shows the same value and
indicate that calibration is in progress.
Resistance Span (2k0)
Set the cell simulator to 2k.
The display advances automatically to the next step once a stable and valid value is
recorded.
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.
Continued on next page.
A:Res.Span(20R)
is displayed to
Calib