...7 CALIBRATION
...7.3 Factory Settings
0. 0
uS/cm
0. 0
Calib
A:Res.Zero (O/C)
500. 0
uS/cm
500. 0
Calib
A:Res.Span (2K0)
0
uS/cm
0
Calib
A:Res.Zero (O/C)
64
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.
Note. The upper 7-segment display shows the measured conductivity. Once the signal is
within range the lower 7-segment display shows the same value and
indicate that calibration is in progress.
Resistance Span (2k0)
Set the cell simulator to 2k.
The display advances automatically to the next step once a stable and valid value is
recorded.
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.