...SPECIFICATION
Access to Functions
Direct keypad access
Measurement, maintenance, configuration, diagnostics or service
functions
Performed without external equipment or internal jumpers
Sensor Cleaning Function – AX416 Only
Configurable cleaning action relay contact
Continuous
Pulse in 1s on and off times
Frequency
5 minutes to 24 hours, programmable in 15 minute increments up
to 1 hour then in 1 hour increments for 1 to 24 hours
Duration
15s to 10 minutes, programmable in 15s increments up to
1 minute then in 1 minute increments up to 10 minutes
Recovery period
30s to 5 minutes, programmable in 30s increments
Mechanical Data
Wall-/Pipe-mount versions
IP66/NEMA4X
Dimensions 192mm high x 230mm wide x 94mm deep
(7.56 in. high x 9.06 in. wide x 3.7 in. deep)
Weight 1kg (2.2 lb)
Panel-mount versions
IP66/NEMA4X (front only)
Dimensions 96mm x 96mm x 162mm deep
(3.78 in. x 3.78 in. x 6.38 in. deep)
Weight 0.6kg (1.32 lb)
Cable Entry Types
Standard
5 or 7 x M20 cable glands
North American
7 x knockouts suitable for
Power Supply
Voltage requirements
85 to 265V AC 50/60 Hz
24V AC or 12 to 30V DC (optional)
Power consumption
<10VA
Insulation
Mains to earth (line to ground) 2kV RMS
72
Environmental Data
Operating temperature limits
–20 to 65C (–4 to 149F)
Storage temperature limits
–25 to 75C (–13 to 167F)
Operating humidity limits
Up to 95%RH non condensing
EMC
Emissions and immunity
Meets requirements of:
EN61326 (for an industrial environment)
EN50081-2
EN50082-2
Hazardous area approvals
CENELEC ATEX IIG EEx n IIC T4
FM non-incendive Class I Div. 2 Groups A to D
CSA non-incendive Class I Div. 2 Groups A to D
Safety
General safety
EN61010-1
Overvoltage Class II on inputs and outputs
Pollution category 2
Languages
Languages configurable:
1
/
in. Hubble gland
2
English
French
German
Italian
Spanish
Pending
Pending
Pending
SS/AX4CO Issue 7