ABB AX410 User Manual page 26

Single and dual input analyzers for low level conductivity
Hide thumbs Also See for AX410:
Table of Contents

Advertisement

...5 PROGRAMMING
...5.3 Configure Sensors
A: T.Comp Range
NH3
set to
Lo TC
NaCl
NaOH
Acid
UPW
-----
Linear
A: Temp.Comp.
24
Low Range Temperature Compensation
Select the type of low range (0 to 100C [32 to 212F]) temperature compensation
required:
Linear
– Linear temperature compensation based on a manually entered
temperature coefficient (see Appendix A1) – see
page 26.
Example
• Non-standard applications.
UPW*
– Temperature compensation based on the temperature coefficient of pure
water. Source data is based on International Standard IEC 60746-3.
Also enables manual entry of a temperature coefficient (see
frame on page 26) for applications where pure water contains an unknown
impurity; in which case the temperature coefficient must be calculated –
see Appendix A1.1.
Acid**
– Temperature compensation based on the temperature coefficient of pure
water containing trace acids.
Examples
• Cation exchanger in-bed and outlet applications.
• Degassed cation/conductivity applications.
NaOH***
– Temperature compensation based on the temperature coefficient of pure
water containing trace caustic.
Example
• Inferred pH in caustic-dosed waters applications.
NaCl*
– Temperature compensation based on the temperature coefficient of pure
water containing trace salts.
Examples
• General monitoring applications.
• Mixed-bed exchanger applications.
• Final polisher effluent applications.
• Cation exchanger inlet applications.
• Anion exchanger in-bed and outlet applications.
• Reverse osmosis applications.
NH3
– Temperature compensation based on the temperature coefficient of pure
**
water containing trace ammonia.
Examples
• Ammonia-treated make-up and boiler feed water applications.
• Condenser sampling applications.
• Hot well sampling applications.
• Before-cation column applications.
• Inferred pH in ammonia-dosed waters applications.
*
Applicable only on conductivities up to 10S cm
** Applicable only on conductivities up to 25S cm
*** Applicable only on conductivities up to 100S cm
A: Temp.Sensor
Continued on page 26.
Temp.Coeff.
Temp.Coeff.
–1
–1
–1
frame on