...4 SETUP
...4.1 Sensor Calibration
A: Calibration
set to
Edit
11. 0 8
mS/cm
1. 0 00
A: Sensor Slope
11. 0 8
mS/cm
0. 0 0
uS/cm
A: Sensor Offset
25. 0
Deg.C
1. 0 00
A: Temp. Slope
25. 0
Deg.C
0. 0
Deg.C
A: Temp. Offset
A: Calibration
set to
Reset
Yes
-----
No
A: Reset?
18
Sensor Slope
Measured conductivity value.
Sensor slope value.
Use the
and
keys to adjust the sensor slope value within the range 0.200 to 5.000
until the measured conductivity value is correct.
Sensor Offset
Measured conductivity value.
Sensor offset value.
Use the
and
keys to adjust the sensor offset value within the range –20.00 to 20.00
until the measured conductivity value is correct.
Temperature Slope
Measured temperature value.
Temperature slope value.
Use the
and
keys to adjust the temperature slope value within the range 0.200 to
1.500 until the measured temperature value is correct.
Temperature Offset
Measured temperature value.
Temperature offset value.
Use the
and
keys to adjust the temperature offset value within the range –40.0 to
40.0C (–40.0 to 104.0F) until the measured temperature value is correct.
Sensor Cal. B
Sensor B calibration (dual input analyzers only) is identical to Sensor A
calibration.
Single input analyzers only – return to top of page.
Sensor Cal. A
Reset Calibration
Select
and press
to reset the calibration data.
Yes
Select
and press
to abort.
No
Return to top of page.