Test Outputs And Maintenance - ABB AX410 User Manual

Single and dual input analyzers for low level conductivity
Hide thumbs Also See for AX410:
Table of Contents

Advertisement

...5.11 Test Outputs and Maintenance

Hold Outputs
set to
Auto.
1
Hrs
30
Mins
Automatic Time
Maintenance
Yes
-----
No
Load/Save Config
Yes
Load
-----
Save
User Config.
Factory Config.
-----
Press
To Set.
Press
To Abort
TEST/MAINTENANCE
Automatic Time
If required, set a time period between 1 and 6 hours, in 30 minute increments, for which the
outputs are held when
Hold Outputs
At the default setting of
None
during sensor calibration and enabled automatically at the end of the procedure.
If a time is set, changes in relay action and analog outputs are disabled during sensor
calibration, but if the calibration is not completed within the set time, the calibration is
aborted, the display returns to the Operating Page and
Continued below.
Load/Save Config
See Section 7.3.
FACTORY SETTINGS
Load/Save Configuration
Select whether a configuration is to be loaded or saved.
Note. If
is selected, pressing the
No
Return to main menu.
TEST/MAINTENANCE
See Section 7.3.
FACTORY SETTINGS
Load User/Factory Configuration
Note. Applicable only if
Load/Save Config
Factory Config.
– resets all the parameters in the
– saves the current configuration into memory.
Save User Config.
Load User Config.
– reads the saved user configuration into memory.
and
User Config.
Factory Config.
saved previously. Use the
and
Press
to Set
Press
Press the appropriate key to load/save the configuration or abort the changes.
is set to
Auto.
, changes in relay action and analog outputs are disabled
CAL. ABORTED
key has no effect.
is set to
.
Yes
Company Standard.
are displayed alternately if a User Configuration has been
and
keys to make the required selection.
are displayed alternately on the lower display line.
to Abort
5 PROGRAMMING
is displayed.
to the
Configuration Pages
49

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax411Ax450Ax455Ax456Ax413Ax416 ... Show all

Table of Contents