...5 PROGRAMMING
...5.5 Configure Outputs
100. 0
uS/cm
0. 0 00
uS/cm
AO1: Span Value
100. 0
uS/cm
0. 0 00
uS/cm
AO1: Zero Value
10. 0 0
uS/cm
12. 0 0
mA
AO1: Set X Value
10. 0 0
uS/cm
12. 0 0
mA
AO1: Set Y Value
36
Span Value
and
are shown alternately on the upper display line. Use the
uS/cm
Adjust
keys to adjust the displayed reading to the required span value. This is Point A in Fig. 5.6.
Zero value.
Zero Value
Span value.
and
are shown alternately on the center display line. Use the
uS/cm
Adjust
keys to adjust the displayed reading to the required zero value. This is Point D in Fig. 5.6.
Note. Applicable only if the
When set to
and
Log. 2
Log. 3
Set Breakpoint X Value
and
are shown alternately on the upper display line. Use the
uS/cm
Adjust
keys to adjust the displayed reading to the required breakpoint conductivity value. This is
Point B in Fig. 5.6.
Current value at which the breakpoint occurs.
Note. Applicable only if the
Set Breakpoint Y Value
Conductivity value at which the breakpoint occurs.
and
are shown alternately on the center display line. Use the
mA
Adjust
to adjust the displayed reading to the required breakpoint current value. This is Point C in
Fig. 5.6.
Note. Applicable only if the
A01: Default O/P
Continued on next page.
parameter is set to
Curve
Linear
, the zero value is set automatically.
parameter is set to
Curve
Bi-Lin
parameter is set to
Curve
Bi-Lin
and
and
or
– see previous page.
Bi-Lin
and
– see previous page.
and
keys
– see previous page.