5.11 Test Outputs and Maintenance

-----
TEST/MAINTENANCE
-----
Test Outputs
4. 0 0
mA
20. 0
%
Test Output 1
-----
Maintenance
Auto.
On
-----
Off
Hold Outputs
48
Test Outputs
Displays the output test details for the analog outputs.
Note. Outputs 3 and 4 are available only if the option board is fitted and analog features
enabled - see section 7.3, page 62.
frame only is shown; the format of frames for the remaining outputs is
Test Output 1
identical.
See below.
Maintenance
Test Output 1
The theoretical output current value.
Output current as a percentage of the full range current.
Use the
to give the output required.
See Section 7.3.
FACTORY SETTINGS
Test remaining outputs.
Test Output 2
Maintenance
Hold Outputs
Enables the relay action and analog outputs to be maintained.
Auto.
On
Off
Note. The LEDs flash while the analyzer is in 'Hold' mode.
Continued on next page.
Load/Save Config
See Section 7.3.
FACTORY SETTINGS
set to
Hold Outputs
Maintenance
set to
Hold Outputs
Automatic Time
and
keys to adjust the displayed theoretical output current value
-
Changes in relay action and analog outputs are disabled during
sensor calibration.
-
Changes in relay action and analog outputs are disabled.
-
Changes in relay action and analog outputs are enabled.
or
- return to main menu.
Off
On
- continued on next page.
Auto.
5 PROGRAMMING