Sensor Stability Relative To Ysi (Accuracy Over Time) - (Adult) - Dexcom G5 User Manual

Hide thumbs Also See for G5:
Table of Contents

Advertisement

Table 8-A. Sensor Stability Relative to YSI (Accuracy Over Time
Day
of
Study
2
Wear
Original
Day
1
Software 505
Original
Day
4
Software 505
Original
Day
7
Software 505
CGM readings are within 40 to 400 mg/dL, inclusive.
1
Both sets of study data are presented and are labeled as
2
505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
Number
Mean
of
Absolute
Paired
Percent
CGM-
Differences
YSI
3023
16.7%
680
10.7%
3108
11.4%
777
8.0%
3021
11.9%
806
8.5%
Median
Percent
Absolute
Within
Percent
15/15%
Differences
YSI
13.2%
59%
7.9%
77%
8.2%
77%
6.4%
89%
8.9%
76%
7.2%
90%
Original
(SW10050) or
) - (Adult)
1
Percent
Percent
Within
Within
20/20%
30/30%
YSI
YSI
71%
86%
84%
96%
87%
95%
96%
99%
87%
95%
97%
99%
Software
Percent
Greater
Than
40/40%
YSI
6%
2%
2%
0%
2%
0%
325

Advertisement

Table of Contents
loading

Table of Contents