Sensor Stability
Sensors can be worn for up to 10 days. Performance was estimated by
calculating the percentage of G6 Pro readings within 15 mg/dL or 15%
(15/15%), 20 mg/dL or 20% (20/20%), and 40 mg/dL or 40% (40/40%),
of the YSI values at the beginning (Day 1, 2), middle (Day 4, 5) and
end (Day 7, 10) of the G6 lifecycle. For blood glucose values less than
or equal to 70 mg/dL, the absolute difference in mg/dL between the
two glucose results was calculated. For values greater than 70 mg/dL,
the absolute difference (%) relative to the YSI values was calculated.
In addition, the mean absolute relative difference (MARD) shows the
average amount the sensor readings differ from the YSI glucose. The
MARD values included in Table 6-A and 6-B show consistent accuracy
and sensor stability over the 10-day life of the sensor.
Table 6-A� Sensor Stability Relative to YSI (Accuracy over Time
(Adults; n=159)
Wear
Period
Beginning
Middle
End
CGM readings are within 40-400 mg/dL, inclusive.
1
G6 Pro User Guide
Appendix G: Technical Information
Number
of
MARD
paired
(%)
CGM-
YSI
6,696
10.9
6,464
9.2
6,169
9.6
Percent
within
15/15% YSI
20/20% YSI
(%)
76.5
84.3
82.3
Percent
Percent
within
within
40/40% YSI
(%)
88.0
94.6
92.4
)
1
(%)
99.6
99.8
99.8
135
Need help?
Do you have a question about the G6 Pro and is the answer not in the manual?
Questions and answers