Server Unavailable Status
Problem
• Share isn't sharing because Dexcom server is offline
Solution
• Wait. Dexcom will fix the issue as soon as possible. For more information, contact
your local Dexcom representative.
Share App
14.10 Accuracy
G6 Reading ≠ Meter Value
Problem
• G6 reading and meter value don't match
Solution
• Different body fluids give different numbers:
• Meter measures glucose from blood.
• Sensor measures glucose from interstitial fluid.
• Calibrating may help align your G6 readings to your meter values.
• See Chapter 7.
Dexcom G6
®
System User Guide
Chapter 14: Troubleshooting
257
Need help?
Do you have a question about the G6 and is the answer not in the manual?
Questions and answers