The Dexcom G5 Mobile System is an M-PED (Medical-Portable Electronic Device), which meets the
FAA RTCA /DO-160 edition G Section 20, Category T and Section section 21, Category M. It can be
used on aircraft according to the directions provided by the operator of the aircraft.
This device can withstand exposure to common electrostatic discharge (ESD) and electromagnetic
interference (EMI).
Guidance and Manufacturer's Declaration –
Electromagnetic Immunity
The transmitter (P/N 9438-06) is intended for use in the electromagnetic environment specified
in the next table. The customer or the user of the transmitter should ensure that it is used in such
an environment.

Transmitter Electromagnetic Immunity Specifications

Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8

Receiver Product Specifications

Part Number
Reading Frequency
Dimensions
TX/RX Frequencies
Bandwidth
Maximum Output Power
Dexcom G5 Mobile System User Guide
Technical Information
IEC 60601 Test
Level
± 8 kV Contact
± 15 kV Air
30 A/m
(Continued on next page)
Transmitter
Electromagnetic
Compliance
Environment Guidance
Level
Floors should be wood, concrete or
± 8 kV Contact
ceramic tile. If floors are covered
with synthetic material, the relative
± 15 kV Air
humidity should be at least 30%.
Magnetic fields from common
30 A/m
appliances are not expected to
affect the system.
MT22949
Every 5 minutes
Length: 4.2 inches
Width: 2.5 inches
Thickness: 0.55 inches
2.402-2.480 GHz
1.22 MHz
2.5 mW EIRP
339