was calculated. The percentages of total readings within 15 mg/dL or 15%, 20 mg/dL or 20%, 30
mg/dL or 30% or greater than 40 mg/dL or 40% are provided in Table 1-A and 1-B. The tables are
categorized within CGM glucose ranges. When you see a CGM reading on your receiver, this table
shows you how likely that reading matches your BG level (measured by YSI in the study).
For example, in the SW10505 Adult Study (Table 1-A), the total number of data pairs considered in the
analysis was 2263. Of these, 93% of the system readings fall within ± 20 mg/dL of the YSI BG values
≤ 80 mg/dL and within ± 20% of YSI BG values > 80 mg/dL.
Dexcom G5 Mobile System User Guide
Technical Information
304