System Agreement To Ysi Within Cgm Glucose Ranges (Adult) - Dexcom G5 User Manual

Hide thumbs Also See for G5:
Table of Contents

Advertisement

Table 1-A. System Agreement to YSI Within CGM Glucose Ranges (Adult)
CGM
Glucose
Study
Range
1
(mg/dL)
Original
Overall
Software 505
Original
40-60
Software 505
Original
61-80
Software 505
Original
81-180
Software 505
Original
181-300
Software 505
Original
301-350
Software 505
Original
351-400
Software 505
CGM readings are within 40-400 mg/dL, inclusive.
1
Both sets of study data are presented and are labeled as
2
505
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
Number
of Paired
2
CGM-YSI
9152
2263
512
120
781
226
3853
738
2784
798
775
229
447
152
Percent
Percent
Within
Within
15/15%
20/20%
YSI
YSI
71%
82%
86%
93%
67%
78%
89%
94%
73%
85%
91%
96%
67%
78%
84%
92%
72%
84%
86%
93%
82%
91%
86%
94%
74%
84%
80%
92%
Original
Percent
Within
30/30%
YSI
92%
98%
88%
98%
94%
99%
91%
98%
93%
98%
97%
98%
91%
97%
Software
(SW10050) or
Percent
Greater
Than
40/40%
YSI
3%
1%
6%
0%
2%
0%
3%
1%
4%
1%
2%
1%
5%
0%
302