Configuring Muse/12Sl - GE CARESCAPE B850 Service Manual

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

CARESCAPE Monitor B850
Citrix Timeout:
Citrix Startup Time:

7.7 Configuring MUSE/12SL

NOTE: MUSE viewing is a licensed software feature. You can configure the MUSE/12SL settings
independent of the license status, but actual viewing of MUSE reports requires that the
MuseView license is enabled.
Settings to send 12SL data
1.
Log in to Webmin.
2.
Select Configuration > MUSE/12SL.
3.
In the MUSE/12SL window, enter the applicable information in the following fields:
-
-
4.
Select Save.
These settings take effect immediately after they are submitted.
Settings to view 12SL data
1.
Log in to Webmin.
2.
Select Configuration > MUSE/12SL.
3.
In the MUSE/12SL window, enter the following:
-
-
-
-
4.
Select Save.
These settings take effect immediately after they are submitted.
76
2106778-001
The number of minutes that a Citrix session will remain open after the
Citrix Window is closed. Keeping the session alive allows users to
quickly return to their last session but locks a Citrix license.
Estimated time, in seconds for connecting Citrix. During this time, the
Citrix Window is displayed as blank, grey rectangle.
Location ID
Identifies the location ID number (within the range 0 to 599)
associated with the patient monitor for searching the MUSE
system.
Site Number
Identifies the site number (within the range 1 to 254) associated
with the patient monitor for searching the MUSE system.
MUSE Web Username
MUSE Web Password
Confirm Password
MUSE Web URL
Username used to authenticate with the MUSE
Web to access 12-lead reports.
Password used to authenticate with the MUSE
Web to access 12-lead reports.
The password and the confirmed password must
be the same.
Used to locate the MUSE Web system to access
12-lead reports. Enter the URL in a valid format.