The following points must be observed during installation:
The meter tube must always be completely full.
The flow direction must correspond to the marking, if there is one.
The maximum torque for all flange connections must be complied with.
The meters must be installed without mechanical tension (torsion, bending).
Wafer design flowmeters with coplanar counter flanges may only be installed with suitable
gaskets.
Only gaskets made from a material which is compatible with the media and its temperature
may be used.
Gaskets must not extend into the flow area since possible turbulence could affect the
accuracy of the meters.
The pipelines must not exert any impermissible forces or torques on the meter.
Do not remove the plugs in the connectors until you are ready to install the electrical cable.
Make sure the gaskets for the housing cover are seated properly. Carefully gasket the cover.
Tighten the cover fittings.
Do not expose the transmitter to direct sunlight and provide for appropriate sun protection
where necessary.
When selecting the installation site, make sure that moisture cannot penetrate the terminal
housing or transmitter compartment.
Make sure the signal cable connectors are plugged in and tightened properly to ensure IP 67
protection.