Parameterization
8.7.4

Error 3/9

Error detection for errors "3" and "9" can be activated/deactivated in the "Error 3/9" menu. The
following settings are possible:
Configuration
"Errors 3+9 off"
(factory setting)
"Both on"
"Error 3 off"
Fig. 47: Current output in the event of an alarm
1 Current output without errors "3" and "9",
output: 20.5 mA (NAMUR NE43)
2 Current output with errors "3" and "9", the
output switches to alarm status (21 ... 23
mA, configurable)
86
Function
If the set measuring range is overshot by more than 3.125 %, the current
output will maintain the 20.5 mA it will have reached by this point until the
flow returns to a level below the upper range value set for the measuring
range.
If the set measuring range is overshot by more than 3.125 %, the current
output will switch to the set fault current (factory setting 22.4 mA). If
QmaxDN is overshot by more than 20 %, error "9" will also be output as a
warning before the device is overloaded.
If the set measuring range is overshot by more than 3.125 %, the current
output will maintain the 20.5 mA it will have reached by this point. If
QmaxDN is overshot by more than 20 %, error "9" will also be output and
the current output will switch to the set fault current.
20,5 mA
20 mA
4 mA
0 %
Qmin
FV4000 / FS4000
100 %
103,125 %
3 Current output with error "9", the output
switches to alarm status at 120 % of
QmaxDN (21 ... 23 mA, configurable)
Qmin Low flow
2
3
1
120 % von
QmaxDN
G00721